Replacement Tube for Philips TOMOSCAN CX and TX CT Systems (#885-00, DA/B 180* 823-02)

DA/B 180*

This 1.5 MHU CT tube is specifically designed as a replacement for use with Philips TOMOSCAN CX* and TOMOSCAN TX* CT systems. The DA 180 uses a housing designed and manufactured by Dunlee. The B 180 is a reload of the original equipment manufacturer’s hou

Tech Data

Product Specifications

Tube Type:
DU 1505, 1.5 MHU (1.06 MJ) insert
Target Diameter:
5 in. (127 mm)
Target Angle:
10°
Power Characteristics:
56/130 kW
Operational Characteristics:
50/60 Hz, 150/180 Hz
Focal Spot:
0.8 / 1.5

Filament Characteristics

Volts:
3.5-11.0 / 6.0-9.5
Amperes:
3.0-5.5 / 3.0-5.5
Rotor Speed:
9600 RPM

Stator Resistance

Main Winding:
14 Ohms
Secondary Winding:
46 Ohms
Maximum Anode Cooling Rate:
240 kHU/min. (2.8 kW)
Shipping Weight:
140 lbs. (63.64 kg)
Warranty:
50,000 slice

Back